Analysing the 6mm LS Mode
The LS mode can be analysed in 4 ways. The first is the traditional
analysis with the mouse where the user is required to specify three
points in the resonance spectrum to be used in the analysis. The
second requires the user to specify the position of the resonance
peak. PRAP will then determine material properties for the minimum
norm. The third automatically locates the resonance peak within a
specified range of frequency and then analyses for the minimum norm.
The fourth uses non-linear regression to directly fit material
properties to the measured resonance spectrum.
As an example, the following is a description of analysing the
thickness extensional resonance of the file LS.PZF included with the
PRAP demonstration. For all types of analysis;
-
Select the Open command from the File menu
Within the Open File dialogue box, go to the
Samples sub directory and load LS.PZF. Note that this is a generated spectrum.
Ensure Thickness Extensional is selected from the
Analysis menu.
From the Window menu, select whether individual
data points are plotted (indicated by open red circles), and whether
data points are joined (by black solid lines). Also select whether
the real part of the graph should be displayed in a linear or
logarithmic scale.
Select the (Rw, Xw)
representation from the Transform menu.
To Analyse with the Mouse
-
Select what is to be determined in the analysis. Select the Analysis
Settings command from the Analysis menu. Select the Options tab.
Using the available check boxes, select one or more of the following options;
Determine material constants
Determine complex circuit constants
Determine Van Dyke's circuit constants
Automatically locate peak
Points to average
The automatic peak location option will select the nearest maximum to
the point clicked in the spectra when selecting the peak to be
analysed. Ensure that 'Determine Parameters for Minimum Norm' is not checked.
-
Holding down the Ctrl key, click on the resonance peak to be analysed
with the left mouse button. This must be in the top part of the
spectrum graph.
The Set Resonance Order dialogue box will appear.
Specify the order of the peak selected. The fundamental peak is of
order 1, etc. Select OK.
Holding down the Ctrl key, click on the lower side
band position with the left mouse button. This should be near the
maximum in the (G/w, B/w)
representation (fs).
Holding down the Ctrl key, click on the upper side
band position with the left mouse button. This should be near the
minimum in the plot of Xw.
To Analyse using Minimization of the Norm
-
Select what is to be determined in the analysis. Select the Analysis
Settings command from the Analysis menu. Select the Options tab.
Using the available check boxes, select one or more of the following options;
Determine material constants
Determine complex circuit constants
Determine Van Dyke's circuit constants
Locate Nearest Peak
Determine Parameters for Minimum Norm
Points to average
The automatic peak location option will select the nearest maximum to
the point clicked in the spectra when selecting the peak to be
analysed. Ensure that 'Determine Parameters for Minimum Norm' is checked.
-
Holding down the Ctrl key, click on the resonance peak to be analysed
with the left mouse button. This must be in the top part of the
spectrum graph.
The Set Resonance Order dialogue box will appear.
Specify the order of the peak selected. The fundamental peak is of
order 1, etc. Select OK.
To Analyse Automatically
-
Select what is to be determined in the analysis. Select the Analysis
Settings command from the Analysis menu. Select the Options tab.
Using the available check boxes, select one or more of the following options;
Determine material constants
Determine complex circuit constants
Determine Van Dyke's circuit constants
Locate Nearest Peak
Determine Parameters for Minimum Norm
Points to average
The automatic peak location option will select the nearest maximum to
the point clicked in the spectra when selecting the peak to be
analysed. Ensure that 'Determine Parameters for Minimum Norm' is checked.
-
While still viewing the Analysis Settings dialogue box, select the
Auto-Analysis tab. Specify the resonance peak order and the range of
frequency within which the peak should be located.
Select the Auto-Analyse command from the Analysis menu.
To Analyse using Non-Linear Regression
-
Select what is to be determined in the analysis. Select the Analysis
Settings command from the Analysis menu. Select the Options tab.
Using the available check boxes, select one or more of the following options;
Determine material constants
Determine complex circuit constants
Determine Van Dyke's circuit constants
Locate Nearest Peak
Determine Parameters for Minimum Norm
Points to average
The last three options have no effect when performing non-linear regression.
-
While still viewing the Analysis Settings dialogue box, select the
Regression tab. Ensure that chi squared is set as desired, and
whether restraints will be used on the fitted properties. If using
restraints, ensure that the range of valid material properties is as desired.
Select the Non-linear Regression command from the
Analysis menu. This invokes the Initial Parameter Values dialogue box
where the initial material properties in the fit are specified, and
the range of frequency over which the resonance spectrum will be
fitted. If the resonance spectra has already been analysed, the
previous results can be used as the initial material properties.
Select Fit.
The progress of the fit will be updated on the
graph. The fitting routine can be aborted at any time. The last best
fit to the spectrum will be used to set results for an aborted fit.
Fitted material properties include an estimate of uncertainty.
When Analysis is Complete
When analysis is complete, the Analysis Results dialogue box appears.
Here the various material properties determined from the length
extensional mode are given.
The resonance spectrum for each selected fit in the Analysis Settings
dialogue box will also be plotted over the original spectrum. The
original data is displayed as open red circles, black solid lines, or
both (o, ---).
The complex material constant fit is displayed as a broken blue line (-----).
The complex circuit constant fit is displayed as a broken green line (-----).
The Van Dyke's circuit model fit is shown as a broken magenta line (-----). |